The short answer
- Start from the part number. Our published models are WS4540-12-NZ03, WS4540-12-NZ01, WS4540-24-NZ01, WS7040AL-12-X200, WS7040AL-24-V200, WS7040-24-V200, WS9250-24-240-X200 and WS8570-24-S300. The number encodes frame size, voltage and variant.
- Five dimensions have to clear. Supply voltage and driver architecture, frame size and footprint, outlet geometry, airflow at the working pressure, and the pressure the device needs.
- Two product pages publish two different static pressure figures. Both are set out below; take the lower as the design limit.
- Published pressure is a zero-flow endpoint. The 6.8 kPa on the WS7040-24-V200 is pressure at zero airflow, and 22 m3/h is flow at zero pressure. Neither is the operating point.
- Replacements are sold at one piece. Our published minimum order quantity is one piece, so a single replacement can be bench-tested before a fleet change.

Why a CPAP blower gets replaced rather than repaired
A blower is a consumable part of the device in a way that a hose or a filter is not, because its failure mode is inside a sealed assembly that cannot be economically opened. The motor, bearing, impeller and scroll housing are integrated at the factory, and published MTTF sits above 20,000 hours under 25 degrees Celsius for most of the range, with one medical variant above 10,000. When that life is consumed, or when bearing noise or delivered pressure tells the engineer the unit is out of specification, the action is a swap.
There is a second reason the swap happens, and it is the reason this guide exists. A CPAP blower is selected for a device operating point, while a replacement ordered from a catalogue is often selected from a headline number. Published maximum pressure and maximum airflow are the two endpoints of one curve and are never delivered together. A device needing 15 L/min at 3 kPa is not served by a unit whose figure is “maximum 5 kPa”, nor by one whose figure is “maximum 8 m3/h”.
The five dimensions a replacement match has to clear
1. Supply voltage and driver architecture. Our published models divide cleanly by supply: 12 VDC for the -12 variants, 24 VDC for every -24 variant. Drive options are PWM, analogue and potentiometer,, some with a built-in driver. A 12 V unit on a 24 V rail fails in seconds; on a 12 V rail a 24 V unit does not spin.
2. Frame size, mounting footprint and outlet geometry. This is where a cross-reference most often fails: the catalogue shows a diameter and a height, not the port positions. Our published sizes run from the 48 mm WS4540 frames at OD12 mm / ID8 mm outlet, through the 60 mm WS7040-24-V200 at D70 x H37 mm with OD17 mm / ID12 mm outlet, to the 90 x 90 x 50 mm WS9250. A replacement matching on flow but not outlet position will not fit the plenum.
3. Airflow at the working pressure. This is the dimension the P-Q curve governs and the one most often skipped. The published flow and pressure pairs are in the table below, and so is the load point at which a model reaches maximum output air power, more useful than either endpoint.
4. The pressure the device actually needs. This is the number the engineer measures on the failed unit rather than looks up. Record the delivered pressure and the flow at it before the unit comes out, because a device that has drifted will be blamed on the new part.
5. Acoustic and thermal margin. A CPAP device is used in a bedroom next to a sleeping person, so noise is a functional requirement, not a comfort preference. We publish the mitigation our customers use: foam or silicone filling between the blower fan and the machine to insulate the noise. The replacement should be the same frame family as the original, because that fix is tuned to the housing.
Cross-reference table: published CPAP blower models
Every figure in this table is a published value for the named model. Work across the row to see which dimension fails first. Maximum flow is the flow at zero pressure, max pressure at zero flow.
| Model | Supply | Max flow at 0 kPa | Max pressure at 0 flow | Published unit size | Weight |
|---|---|---|---|---|---|
| WS4540-12-NZ03 | 12 VDC | 7.2 m3/h | 5.0 kPa | 48 mm frame | 63 g |
| WS4540-12-NZ01 | 12 VDC | 8 m3/h | 4.8 kPa | 48 mm frame | 63 g |
| WS4540-24-NZ01 | 24 VDC | 7.5 m3/h | 4.8 kPa | 48 mm frame | 63 g |
| WS7040AL-12-X200 | 12 VDC | 19 m3/h | 5.5 kPa | 92 mm frame | not published |
| WS7040AL-24-V200 | 24 VDC | 16 m3/h | 6.5 kPa | D60 x H40 mm | 134 g |
| WS7040-24-V200 | 24 VDC | 22 m3/h | 6.8 kPa | D70 x H37 mm | 80 g |
| WS9250-24-240-X200 | 24 VDC | 44 m3/h | 8 kPa | 90 x 90 x 50 mm | 400 g |
| WS8570-24-S300 | 24 VDC | 47 m3/h | 15.0 kPa | D89 x H70 mm | 500 g |
Two notes on reading that table. Frame size is not a proxy for performance: two 48 mm WS4540 models differ by 0.8 m3/h of flow and 0.2 kPa. And the WS7040-24-V200 and WS7040AL-24-V200 sit close on paper yet are different frames at 80 g and 134 g, so they are not interchangeable.
For the full range we publish for respiratory applications, start from the CPAP replacement blower category page.
Reading the published data where two figures disagree
A guide that silently picks one of two conflicting published numbers is worse than one that flags the conflict, because the engineer will not know which number they trust. Two models above have this issue today.
On the WS7040AL-24-V200 page, the specification block lists static pressure as 7.6 kPa, the descriptive paragraph states 6.5 kPa, and the heading line calls it a 60 mm 7 kPa blower: three numbers for one model. On the WS8570-24-S300 page, the block and paragraph agree at 15.0 kPa, but the title line calls it a 70 mm 7 kPa blower. There the conflict sits between the technical block and the title line.
The rule does not depend on resolving which figure is correct: take the lower published figure as the design limit and treat the higher one as unverified until a datasheet says otherwise. The WS7040AL-24-V200 is then a 6.5 kPa part and the WS8570-24-S300 a 7 kPa part, the reading that will not produce a device losing pressure.
One more published field decides most replacements on cost of ownership rather than fit. Published MTTF is more than 20,000 hours under 25 degrees Celsius for most of this range and more than 10,000 for the WS7040AL-24-V200. That spread is a reason to prefer one model in a fleet programme, and to check the ambient temperature.
How to order a CPAP blower replacement that arrives the first time
Most replacement orders fail on information, not hardware. The sequence below produces a unit that drops in first time.
- Record the failed part number and the nameplate voltage before ordering. The part number gives the frame family and the variant; the nameplate gives the rail the device uses. If the nameplate is unreadable, measure at the blower connector under the device’s own supply.
- Measure the delivered pressure and flow of the failed unit, if it will still run. A replacement selected against a catalogue maximum rather than the measured duty point will over-deliver, showing up as noise and power draw, or under-deliver, showing up as a complaint.
- Take the frame size and outlet geometry from the failed unit with calipers, not the catalogue. Measure the mounting hole pattern, outlet diameter and position, since that position varies by model.
- Establish the drive configuration before the order. Some models carry a built-in driver while others need an external controller board, and a unit with a built-in driver driven by an external board will not behave as specified.
- Confirm the connector and the pinout, and treat it as a separate line item. The harnesses terminate in two white connectors with different pin counts, and lead lengths differ between frames. A blower that is electrically and mechanically correct can still be unusable because the connector does not reach.
- Ask for the acoustic position in writing. Published noise figures are model-specific, and the foam or silicone filling between fan and machine is tuned to a housing, so specify that frame.
- Request the conformance evidence that applies to your market. Published certification positions are CE and RoHS for most of this range, with ETL additionally on four models and ISO 9001 on the WS8570-24-S300. For a device sold in the United States the equipment standard is IEC 60601-1-2 on electromagnetic disturbances, and the requirements for sleep apnoea equipment are set out in ISO 80601-2-70. To match a model to a specification, find your replacement model and send the part number and measured duty point.
Frequently asked questions about CPAP blower replacement
Can I replace a CPAP blower with any blower of the same diameter?
No. Frame size is one of five dimensions that have to clear,, the least decisive. Two published 48 mm WS4540 models differ by 0.8 m3/h of airflow and 0.2 kPa, and the WS7040-24-V200 and WS7040AL-24-V200 differ by 6 m3/h and 54 g. Diameter alone will not tell you whether the part fits the plenum, or holds pressure. Match on the part number first, then confirm voltage, footprint, outlet and duty point.
What is the difference between a CPAP blower and a BiPAP or ventilator blower?
They are related but not the same part. Our CPAP category covers the 48 mm, 60 mm and 92 mm frames listed above, and the same product pages name air cushion machines, fuel cell and medical equipment alongside CPAP as published applications, so a shared frame can serve several markets. A BiPAP or ventilator device has a different flow and pressure requirement, and the standard differs: ISO 80601-2-70 applies to sleep apnoea equipment and excludes equipment for patients dependent on artificial ventilation. If the device is a ventilator, do not use a CPAP part number as the cross-reference key.
The published maximum pressure and the published maximum airflow do not match my device. Which one is right?
Neither, on its own, and this is the most common misreading in a cross-reference. Both are endpoints of the same P-Q curve: maximum airflow at zero pressure, maximum pressure at zero flow. Your device needs a point in between, so neither can be matched directly. We publish the load points at which a model reaches maximum output air power and maximum efficiency, the useful reference points for a drive setting.
What do I do when the product page gives two different static pressure figures?
Take the lower figure as the design limit and treat the higher one as unverified until a part-number-level datasheet says otherwise. Two published models currently show this pattern, described in full above. Under that rule the WS7040AL-24-V200 is a 6.5 kPa part and the WS8570-24-S300 a 7 kPa part. Ask for the figure that belongs to your exact part number and revision, in writing, on a datasheet.
How many can I order, and how much notice should I allow?
Our published minimum order quantity for this range is one piece,, with samples orderable for testing before a larger order. The workflow is to buy one unit, fit it and confirm the duty point on the bench, then commit to a fleet quantity. That avoids buying a batch against a catalogue number that differs from the fitted part in one of the five dimensions above.
Does replacing the blower change the device’s compliance status?
It can, and that is a change you control rather than the distributor. A blower is part of the finished device, so swapping it changes the acoustic output and electrical behaviour of a certified device. Our published certification positions are CE and RoHS across most models, with ETL additionally on four models and ISO 9001 on the WS8570-24-S300. Those are component positions; the device-level obligation sits with you.
Where to go from here
If you have the failed part number and the measured duty point, that is enough to identify a replacement. Send the part number, supply voltage, measured pressure and flow, and frame size with port dimensions, and we will return the published specification position in writing, with the conformance position.
Reference pages: the mini respirator ventilator CPAP blower page for the WS7040-24-V200, and the find your replacement model contact page. For clinical questions outside component specification, the Resmed FAQs cover CPAP operation.
About the author
Mr. Dong, Technical Director at Ningbo Wonsmart Motor Fan Co., Ltd.
Mr. Dong serves as the Technical Director at Ningbo Wonsmart Motor Fan Co., Ltd., leading the company’s brushless DC blower engineering and application support across medical, industrial, and laser equipment sectors. He works closely with OEM customers on motor specification alignment, integration fit verification, and compliance documentation for IEC 60601-1 medical devices and industrial EMC / RoHS / REACH standards.
Sources. All model numbers, supply voltages, airflow and static pressure figures, unit sizes, outlet diameters, weights, certification positions and MTTF values here are manufacturer published values of Ningbo Wonsmart Motor Fan Co., Ltd., from its CPAP blower category page and the product pages for the eight models in the table. Where a page publishes two static pressure figures for one model, both are reported here and the lower figure is used. No unit prices are published and none are quoted. No market statistics or survey data are cited.
Post time: Nov-19-2026